Magnetization
The magnetization in pulsed fields is measured by integrating the voltage induced in a compensated coil system surrounding the sample (induction method) [1].
The technique is best suited for magnet type D (on request can be installed in magnets A, B, E).
Standard temperature range is 1.4 – 270 K.
3He option with a base temperature of ~0.6 K is available for magnets A, B, E.
Pickup-coil system used in the pulsed-field magnetometer with (left) the principal sketch, (middle) the electrical scheme, and (right) a picture of the original set up.
Measurement example: Magnetization M and its field derivative dM/dH of an Ising-like spin-1/2 antiferromagnetic chain system BaCo2V2O8 as function of magnetic field.
[1] Y. Skourski et al., Physical Review B 83 214420 (2011) (1)
[2] Z. Wang et al. Phys. Rev. Lett. 120 (2018), 207205 (2)
URL of this article
https://www.hzdr.de/db/Cms?pOid=45932
Links of the content
(1) | http://journals.aps.org/prb/abstract/10.1103/PhysRevB.83.214420 |
(2) | https://journals.aps.org/prl/pdf/10.1103/PhysRevLett.120.207205 |