Technical feasibility of integrating 7 T anatomical MRI in image‑guided radiotherapy of glioblastoma: a preparatory study


Technical feasibility of integrating 7 T anatomical MRI in image‑guided radiotherapy of glioblastoma: a preparatory study

Compter, I.; Peerlings, J.; Eekers, D. B. P.; Postma, A. A.; Ivanov, D.; Wiggins, C. J.; Kubben, P.; Küsters, B.; Wesseling, P.; Ackermans, L.; Schijns, O. E. M. G.; Lambin, P.; Hoffmann, A. L.

Objectives: The use of 7 Tesla (T) magnetic resonance imaging (MRI) has recently shown great potential for high-resolution soft-tissue neuroimaging and visualization of microvascularization in glioblastoma (GBM). We have designed a clinical trial to explore the value of 7 T MRI in radiation treatment of GBM. For this aim we performed a preparatory study to investigate the technical feasibility of incorporating 7 T MR images into the neurosurgical navigation and radiotherapy treatment planning (RTP) systems via qualitative and quantitative assessment of the image quality.
Materials and methods: The MR images were acquired with a Siemens Magnetom 7 T whole-body scanner and a Nova Medical 32-channel head coil. The 7 T MRI pulse sequences included magnetization-prepared two rapid acquisition gradient echoes (MP2RAGE), T2-SPACE, SPACE-FLAIR and gradient echo sequences (GRE). A pilot study with three healthy volunteers and an anthropomorphic 3D phantom was used to assess image quality and geometrical image accuracy.
Results: The MRI scans were well tolerated by the volunteers. Susceptibility artefacts were observed in both the cortex and subcortical white matter at close proximity to air-tissue interfaces. Regional loss of signal and contrast could be minimized by the use of dielectric pads. Image transfer and processing did not degrade image quality. The system-related spatial uncertainty of geometrical distortion-corrected MP2RAGE pulse sequences was ≤2 mm.
Conclusion: Integration of high-quality and geometrically-reliable 7 T MR images into neurosurgical navigation and RTP software is technically feasible and safe.

Keywords: Ultra-high field MRI; Radiotherapy; Treatment planning; Glioblastoma; Geometrical distortion

Permalink: https://www.hzdr.de/publications/Publ-24783