TSEE Dosimetry of Soft X-rays
J. Pawelke, A. Panteleeva

The measurement of RBE by cell irradiation requires precise determination of dose, delivered to the cell target. Absolute dose measurements are usually done by an air-filled ionisation chamber. However, in the case of low energy X-rays (below 10 keV) it does not sample the cell target because of the high attenuation. Detection based on thermally stimulated exoelectron emission (TSEE) allows to sample the depth dose distribution in a cell monolayer ( »  5 mm thick), because the thickness of the sensitive detector layer is of the order of ~  10 nm [1]. A TSEE prototype system (Dr. Holzapfel Messgerätelabor, Teltow) consisting of 3 types of thin-film BeO (100 nm) detectors and a counter based on a spherical anode Geiger-Müller methane gas-flow counter was tested. The irradiation was performed with a high-activity beta-source (90Sr), a calibrated gamma-source (137Cs) and a soft X-ray source (55Fe). Highest response as well as stability were shown by detector type I (fig. 1, table 1, same results were obtained with 55Fe source), which proves that the 200 nm Au layer contributes for better mechanical attachment of the BeO layer as well as for electrical stability. The saturation in the dose response (fig. 2) due to the limited pulse resolution of the gas-flow counter results in an upper detection limit of about 0.1 Gy and is far from the requirements of the cell survival studies ( £  10 Gy). Therefore it cannot be used as online monitoring system. In order to determine the dose delivered to cells, measurements must be done in the presence of cell culture medium. Irradiations were made after wetting the surface and showed no change in the detector response due to any tribo-effect [2]. As expected detector response after irradiation of detectors covered with about 0.5 mm propanol with low energy X-rays from 55Fe source and high energy photons from 22Na source was reduced only due to attenuation in the liquid layer. However, the dependence of the detector response on photon energy and dose rate must still be studied.

pawelke1.gif pawelke2.gif
Fig. 1 Typical glow curves of thin-film BeO detectors. Fig. 2 TSEE response of detector type I to electron (90Sr) and photon (137Cs) irradiation.



Table 1 Main detector properties.

Detector Layer Sensitive Peak range Reproducibilitya Sensitivityb
type structure area [mm2] [oC] [%] [counts/mGy]
I BeO-Au-C   43.0 180...360 1.8 15538
II BeO-C-Al 146.4 200...410 2.6   7603
III BeO-C 146.4 180...360 9.6   6137

 a) average error of the mean for a series of 8 measurements with 90Sr      b) 137Cs, dose in air   

References

[1] J.S. Durham et. al., Radiation Protection Dosimetry, Vol. 39(1991), No 1/3, pp. 67-70
[2] H. Eichenüller, Radiation Protection Dosimetry, Vol. 4(1983), No 3/4, pp. 281-285

FZR
 IKH 06/27/01 © J. Pawelke