Publications Repository - Helmholtz-Zentrum Dresden-Rossendorf
1 PublicationSource term calculation and validation for F-18 production with a cyclotron for medical applications at HZDR
Konheiser, J.; Müller, S. E.; Magin, A.; Naumann, B.; Ferrari, A.
In this document we present the calculation and experimental validation of a source term for F-18 production with a cyclotron for medical applications operating at 18 MeV proton energy and 30 microA proton current. The Monte Carlo codes MCNP6 and FLUKA were used for the calculation of the source term. In addition, the radiation field around the O-18 enriched water target was simulated with the two codes. To validate the radiation field obtained in the simulation, an experimental program has been started using activation samples which are placed close to the water target during an F-18 production run of the cyclotron. After the irradiation, the samples are analyzed and the resulting activation is compared to Monte Carlo calculations of the expected sample activation. We and good agreement between simulations and experimental results, with most calculation to experiment (C/E) ratios well between 0.6 and 1.4.
Keywords: F-18; MCNP; FLUKA; Source Term; Cyclotron
Involved research facilities
- PET-Center
Related publications
-
Source term calculation and validation for F-18 production with a cyclotron …
ROBIS: 28964 HZDR-primary research data are used by this (Id 28971) publication
-
Journal of Radiological Protection 39(2019), 906-919
Online First (2019) DOI: 10.1088/1361-6498/ab2ae8
Cited 5 times in Scopus
Permalink: https://www.hzdr.de/publications/Publ-28971