Publications Repository - Helmholtz-Zentrum Dresden-Rossendorf
1 PublicationQualification of Image–Based Measurement Systems for Characterization of Sprays
Schulz, J.; Schunk, C.; Schleicher, E.; Bart, H.-J.
Image-based measurement techniques become increasingly popular and expedite digitalization in chemical engineering. This article demonstrates their potential by testing two inline probes, namely modified optical multimode online probe (OMOP) and process microscope. Validations are performed with static monodisperse standards (9.2 µm to 406 µm) and fast-moving droplets (68.6 µm to 860.7 µm; 24.5 m s−1 to 11 m s−1). Screening of a lithography attests both probes great distortion-free image quality. A 1951 USAF chart attests a low optical resolution of 8 µm or 7 µm with respect to the OMOP or process microscope, respectively. The modified OMOP and process microscope reaches accuracies of 7.6 % or 5.9 % for particles and 8.2 % or 6.8 % for droplets.
Involved research facilities
- TOPFLOW Facility
Related publications
-
Qualification of Image-Based Measurement Systems for Characterization of Sprays
ROBIS: 36460 HZDR-primary research data are used by this (Id 36459) publication -
Qualification of Image-Based Measurement Systems for Characterization of Sprays
RODARE: 2120 HZDR-primary research data are used by this (Id 36459) publication
-
Chemie Ingenieur Technik 93(2021)7, 1111-1118
DOI: 10.1002/cite.202000104
Cited 1 times in Scopus
Permalink: https://www.hzdr.de/publications/Publ-36459