Void measurement using high-resolution gamma-ray computed tomography


Void measurement using high-resolution gamma-ray computed tomography

Bieberle, A.; Hoppe, D.; Schleicher, E.; Hampel, U.

We present a high-resolution gamma-ray computed tomography (CT) system for the measurement of cross-sectional time-averaged void distributions in thermo hydraulic facilities. The system has been carefully designed for harsh operating conditions, such as varying temperature fields and strong magnetic fields, typically produced by thermal hydraulic test loops with direct electric bundle heating. Measurements are non invasive, thus the two-phase flow in the test section is not influenced. The gamma-ray CT system consists of a collimated 137Cs isotopic source, a gamma radiation detector arc including 320 single elements, a pulse processing unit and a thermal stabilisation unit. The spatial resolution of the CT system is about 2 mm in plane. Recently, the thermal design of the detector arc is improved to secure maintenance of constant temperature of thermally sensitive components under changing environmental conditions. This turned out to be a key issue for achieving accurate quantitative measurements. First results of laboratory measurements on a bundle mock-up with this improved system are presented.

Keywords: Gamma-ray computed tomography