Measuring the Saturation Magnetization in Samples with Unknown Magnetic Volume due to Intermixing by Ion Irradiation


Measuring the Saturation Magnetization in Samples with Unknown Magnetic Volume due to Intermixing by Ion Irradiation

Markó, D.; Lenz, K.; Strache, T.; Kaltofen, R.; Fassbender, J.

We present a method to determine the saturation magnetization of samples for which the magnetic volume is unknown and thus cannot be calculated from the magnetic moment. This can happen, e.g., in multilayers, where the spacer material is likely to cause intermixing or whenever ion irradiation is used to modify the magnetic properties of samples on purpose. In both cases the active magnetic volume is altered from its nominal value in unknown manner. Therefore magnetometry like supraconducting quantum interference devices (SQUID) or vibrating sample magnetometry (VSM) fail, because they detect the magnetic moment but do not provide information on the respective magnetic volume. In this article we have used thin films of Permalloy (Py) and Py/Ta multilayers. Some of the Py/Ta samples were irradiated with Ne ions in order to modify the interfacial mixing. The saturation magnetization is determined by ferromagnetic resonance (FMR).

Keywords: Ferromagnetic Resonance; Saturation Magnetization; Ion Irradiation; Intermixing

  • Contribution to proceedings
    MMM/Intermag, 18.-22.01.2010, Washington DC, USA
    IEEE Transactions on Magnetics 46(2010)6, 1711-1714

Permalink: https://www.hzdr.de/publications/Publ-13315