Comparison of pancreatic respiratory motion management with three abdominal corsets for particle radiation therapy


Comparison of pancreatic respiratory motion management with three abdominal corsets for particle radiation therapy

Dolde, K.; Schneider, S.; Stefanowicz, S.; Alimusaj, M.; Flügel, B.; Saito, N.; Troost, E. G. C.; Pfaffenberger, A.; Hoffmann, A. L.

Background and Purpose: Abdominal organ motion seriously compromises the targeting accuracy for particle therapy in patients with pancreatic adenocarcinoma. This study compares three different abdominal corsets regarding their ability to reduce pancreatic motion and their potential usability in particle therapy.

Materials and Methods: A patient-individualized polyurethane (PU), a semi-individualized polyethylene (PE), and a patient-individualized 3D-scan based polyethylene (3DPE) corset were manufactured for one healthy volunteer. Time-resolved volumetric magnetic resonance imaging (4D-MRI) and single-slice 2D cine-MRI scans were acquired on two consecutive days to compare free-breathing motion patterns with and without corsets. The corset material properties, such as thickness variance, material homogeneity in Hounsfield units (HU) on CT scans, and manufacturing features were compared. The water equivalent ratio (WER) of corset material samples was measured using a multi-layer ionization chamber for proton energies of 150 MeV and 200 MeV.

Results: All corsets reduced the pancreatic motion on average by 9.6 mm in inferior-superior and by 3.2 mm in anterior-posterior direction. With corset, the breathing frequency was approximately doubled and the day-to-day motion variations were reduced. The WER measurements showed an average value of 0.993 and 0.956 for the PE and 3DPE corset, respectively, and of 0.298 for the PU corset. The PE and 3DPE corsets showed a constant thickness of 2.8 ± 0.2 mm and 3.8 ± 0.2 mm, respectively and a homogeneous material composition with a standard deviation (SD) of 31 HU and 32 HU, respectively. The PU corset showed a variable thickness of 4.2−25.6 mm and a heterogeneous structure with air inclusions with an SD of 113 HU.

Conclusion: Abdominal corsets are effective devices to reduce pancreatic motion. For particle therapy, PE-based corsets are preferred over PU-based corset due to their material homogeneity and constant thickness.

Keywords: Motion management; pancreatic cancer; abdominal corset; image-guided radiotherapy; 4DMRI; particle therapy