Fully characterised and online monitored beamline for high dose rate laser proton irradiation experiments at Draco PW


Fully characterised and online monitored beamline for high dose rate laser proton irradiation experiments at Draco PW

Brack, F.-E.; Kroll, F.; Gaus, L.; Bernert, C.; Beyreuther, E.; Cowan, T.; Karsch, L.; Kraft, S.; Leßmann, E.; Metzkes-Ng, J.; Pawelke, J.; Rehwald, M.; Reimold, M.; Schlenvoigt, H.-P.; Schramm, U.; Sobiella, M.; Umlandt, M. E. P.; Ziegler, T.; Zeil, K.

Laser-driven proton pulse provide unique properties in terms of pulse structure (ns) and instantaneous dose rate (10^9 Gy/s) but - inherently broadband and highly divergent - pose a challenge to established beamline concepts on the path to application-adapted irradiation field formation, particularly for three-dimensional cases. We present the successful implementation and characterisation of a highly efficient and tuneable dual pulsed solenoid beamline at the Draco PW facility[1] to generate volumetric dose distribution tailored to specific applications[2].
The vast experimental scope and already successfully performed studies range from systematic volumetric in-vivo tumour irradiations in a dedicated mouse model (with a stable mean dose delivery of ±10 % and a spatial dose homogeneity of ±5 % over a cylindrical volume of 5 mm diameter and height) to high-dose-rate irradiations in the FLASH regime (using proton peak dose rates of up to 10^9 Gy/s with about 20 Gy/shot homogeneously over a cylindrical sample volume of 4.5 mm diameter and 3 mm height) as well as particle diagnostics commissioning (with a multitude of spatial and spectral dose distributions).
The beamline setup is complemented by a complex beam monitoring and dosimetry detector suite adapted to the ultra-high dose rate pulses and is in its unique synergy and redundancy capable of %-level precision dose delivery to samples as required for systematic irradiation studies. In addition to established radiochromic film dosimetry, the detector suite includes saturation-corrected (transmission) ionisation chambers [3] as well as screen and bulk scintillator setups, partly with tomographic reconstruction capabilities for 3D dose distribution retrieval. Moreover, non-invasive, single-shot-capable online time-of-flight-based spectral characterisation of filtered proton pulses has proven a powerful tool for beam monitoring as well as dosimetric purposes.
In this presentation the complex and versatile dose delivery system of laser-driven protons at the Draco PW using pulsed solenoids will be discussed. Its characterisation, technological development and improvement as well as the dosimetry suite as a vital part of the precise dose delivery will be addressed, while the presentation by U. Schramm covers recent experimental activities in detail.
[1] T. Ziegler, et al., Proton beam quality enhancement by spectral phase control of a PW-class laser system, https://arxiv.org/abs/2007.11499 (2020)
[2] Brack, et al., Spectral and spatial shaping of laser-driven proton beams using a pulsed high-field magnet beamline, SciRep, 10:9118, (2020)
[3] Gotz M, et al., A new model for volume recombination in plane‐parallel chambers in pulsed fields of high dose‐per‐pulse. Phys Med Biol., 62: 8634, (2017)

  • Invited lecture (Conferences) (Online presentation)
    SPIE 2021 ALPA, 21.04.2021, Prague, Czech Republic
  • Lecture (others) (Online presentation)
    LPA online seminar, 10.03.2021, Dresden, Germany

Permalink: https://www.hzdr.de/publications/Publ-33639