Attenuation and Scatter Correction in Fully 3D Limited Angle PET


Attenuation and Scatter Correction in Fully 3D Limited Angle PET

Pönisch, F.; Enghardt, W.; Lauckner, K.

The in-beam dual head positron camera BASTEI (Beta+ Activity meaSurements at the Therapy with Energetic Ions) is used to monitor and control the applied dose distributions simultaneously to tumor irradiations with carbon ion beams at the experimental heavy ion therapy facility at GSI Darmstadt. Therefore, the PET system has been mounted directly at the treatment site. A fully 3D reconstruction algorithm based on the Maximum Likelihood Expectation Maximization algorithm has been developed and adapted to a spatially varying imaging situation. The scatter and attenuation correction is applied to the measured list mode data before each iterative step. This requires an attenuation map containing the information on the tissue composition and densities. This information is derived from the X-ray computed tomograms (CT) of the patient and the patient fixation system including the head rest. The scatter correction is included into the forward projection step of the Maximum Likelihood image reconstruction. The normalization of scattered events relative to the unscattered events is done by a global scatter fraction factor. The results are presented.

  • Poster
    Proc. IEEE Nuclear Science Symposium and Medical Imaging Conference, San Diego, USA (2001) M13B-23
  • Contribution to proceedings
    Proc. IEEE Nuclear Science Symposium and Medical Imaging Conference, San Diego, USA (2001) M13B-23

Permalink: https://www.hzdr.de/publications/Publ-5532