2 PublicationsA Method for the Intrinsic Calibration of CsI(Tl) Detectors
Kamanin, D. V.; Wagner, W.; Ortlepp, H.-G.
Abstract
Abstract
A method for the intrinsic energy calibration of photomultiplier-coupled CsI(Tl) detectors is described. A simple empirical model of the scintillation light pulse-shape of CsI(Tl) crystals for light charged particles has been applied to simulate the particle identification matrix as it follows from the pulse-shape analysis method. The calibration procedure for the large-area CsI(Tl) detectors of the scintillator shell of the 4 π - array FOBOS for ions with Z < 4 at energies below 100 AMeV is based on the energies of the particle punch-through points.
-
Forschungszentrum Rossendorf; FZR-197
ISSN: 1436-3976 -
Nuclear Instruments and Methods in Physics Research A 413 (1998) 127-137
DOI: 10.1016/S0168-9002(98)00467-7
Cited 5 times in Scopus