Publications Repository - Helmholtz-Zentrum Dresden-Rossendorf

1 Publication

Qualification of Image–Based Measurement Systems for Characterization of Sprays

Schulz, J.; Schunk, C.; Schleicher, E.; Bart, H.-J.

Abstract

Image-based measurement techniques become increasingly popular and expedite digitalization in chemical engineering. This article demonstrates their potential by testing two inline probes, namely modified optical multimode online probe (OMOP) and process microscope. Validations are performed with static monodisperse standards (9.2 µm to 406 µm) and fast-moving droplets (68.6 µm to 860.7 µm; 24.5 m s−1 to 11 m s−1). Screening of a lithography attests both probes great distortion-free image quality. A 1951 USAF chart attests a low optical resolution of 8 µm or 7 µm with respect to the OMOP or process microscope, respectively. The modified OMOP and process microscope reaches accuracies of 7.6 % or 5.9 % for particles and 8.2 % or 6.8 % for droplets.

Involved research facilities

  • TOPFLOW Facility

Related publications

Permalink: https://www.hzdr.de/publications/Publ-36459


Years: 2023 2022 2021 2020 2019 2018 2017 2016 2015