Strahlfänger
ELBE uses a linear accelerator without recirculation for a wide variety of applications. This, as well as the arrangement of the experiments in individual radiation-shielded caves requires the installation of several beam dumps with different performance requirements. The following table summarizes the design parameters of the individual beam dumps.
Most ELBE beam dumps share a common design, shown
in the picture below. The beam is stopped in a graphite block held in vacuum
which dissipates the generated heat by radiation only. The dump vacuum is
separated from the beamline UHV by a 0.5mm thick Be window. The stainless-steel
vessel of the dump and the window are water cooled. Every beam dump has its
individual (primary) cooling circuit connected via heat-exchanger to the
main cooling system. The primary circuit containes enough water to accept
the stored energy of the graphite block in case of failure of the main cooling
circuit. The pumps of the primary coolant circuit are operated with a power
backup. The beamdump vessel is surrounded by a 10-20 cm iron radiation shielding,
additional lead shields are installed as required by the particular location. |